Abstract

BackgroundTo investigate whether patient-specific instrumentation (PSI) improves the femoral component positioning of implants during unicompartmental knee arthroplasty (UKA) using cadaver bone models.MethodsFifty adult cadaveric femoral bone specimens collected from February 2016–2018, were randomized to receive medial UKA with a PSI guide (n = 25) or conventional instrumentation (CI) (n = 25). Standard anteroposterior and lateral view radiographs were obtained postoperatively to assess the coronal and sagittal positioning of the femoral prostheses, respectively. The osteotomy time was recorded to assess the convenience of PSI in guiding osteotomy.ResultsOsteotomy time significantly shortened in the PSI group (3.12 ± 0.65 versus 4.33 ± 0.73 min, p < 0.001). There was a significant difference in the postoperative coronal alignment of the femoral component between the PSI and CI groups (varus/valgus angle: 1.43 ± 0.93° vs. 2.65 ± 1.50°, p = 0.001). The prevalence of outliers in coronal alignment was lower in the PSI than the CI group (2/25, 8% vs. 9/25, 36%). Sagittal posterior slope angle of the femoral component was significantly different between the two groups (8.80 ± 0.65° and 6.29 ± 1.88° in the CI and PSI groups, respectively, p < 0.001). The malalignment rate of the femoral component in the sagittal plane was 60% in the CI group, whereas no positioning deviation was observed in the PSI group.ConclusionThis study used a cadaver model to support the fact that CT-based PSI shows an advantage over CI in optimizing implant positioning for UKAs.