Abstract
The initial stability and long-term survivorship of cementless acetabular components is partially dependent on bone-prosthesis contact patterns. This study addresses the question: What are the differences in contact patterns for hemispheric, dual geometry, and spiked designs? Quantitative and qualitative assessment of contact patterns was achieved using a novel, nondestructive computed tomography-based analysis. Eighteen cadaveric hemipelvis specimens were randomly assigned to receive one of three cup designs. The mean amount of cup contact was 48.6% in the hemisphere group, 31.9% in the dual geometry group, and 35.1 percent in the spiked group. The hemispheric design showed more contact and a more uniform contact pattern than the dual geometry design. The dual geometry design featured a tight fit at the rim of the implant. In the spiked group, the amount of bone-prosthesis contact was affected by the distance that the spikes advanced into the pelvic bone. This nondestructive imaging technique has potential future uses in in vitro and in vivo studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.