Abstract

The aim of this study was to create AO 44A1, 44B1, and 44C1 fractures using finite element analysis to determine the stability of Kirschner wire, intramedullary screw, and plate-screw fixation methods in fracture. Using finite element analysis, the postreduction behavior of AO 44A1, 44B1, and 44C1 fractures with Kirschner wire, intramedullary screw, and plate-screw fixation methods was analyzed and compared in terms of displacement and stress. The lowest amount of displacement was provided with the intramedullary screw method in AO 44A1 and 44B1 fractures and with the 4-mm Kirschner wire method in AO 44C1 fractures. The total displacement of the intramedullary screw system used for fixation in AO 44A1, 44B1, and 44C1 fractures was lower. According to finite element analysis results, the lowest amount of displacement was obtained with intramedullary screw fixation in AO 44A1 and 44B1 fractures, and 4-mm Kirschner wire fixation was achieved in AO 44C1 fractures.