Abstract

AbstractSoft robotic devices containing multiple actuating elements have successfully recapitulated complex biological motion, leading to their utility in biomedical applications. However, there are inherent nonlinear mechanics associated with soft composite materials where soft actuators are embedded in elastomeric matrices. Predicting their overall behavior prior to fabrication and subsequent experimental characterization can therefore present a hurdle in the design process and in efficiently satisfying functional requirements and specifications. In this work, a computational design framework for optimizing the motion and function of biomimetic soft robotic composites is demonstrated by conducting a design case study of soft robotic cardiac muscle (myocardium) with a particular focus on applications including replicating and assisting cardiac motion and function. A finite element model of a soft robotic myocardium is built, in which actuators are prescribed with anisotropic strain to simulate local deformation, and various design parameters are investigated by evaluating the performance of each configuration in terms of ventricular twist, volumetric output, and pressure generation. Then, an optimized design is proposed that recapitulates the physiological motion and hemodynamics of the heart, and its thrombogenicity is further explored using a fluid‐structure interaction model. This framework has broader utility in predicting the performance of other soft robotic embedded composites.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call