Abstract

We developed a computational model to investigate the hemodynamic effects of a pulsatile left ventricular assist device (LVAD) on the cardiovascular system. The model consisted of 16 compartments for the cardiovascular system, including coronary circulation and LVAD, and autonomic nervous system control. A failed heart was modeled by decreasing the end-systolic elastance of the ventricle and blocking the mechanism controlling heart contractility. We assessed the physiological effect of the LVAD on the cardiovascular system for three types of LVAD flow: co-pulsation, counter-pulsation, and continuous flow modes. The results indicated that the pulsatile LVAD with counter-pulsation mode gave the most physiological coronary blood perfusion. In addition, the counter-pulsation mode resulted in a lower peak pressure of the left ventricle than the other modes, aiding cardiac recovery by reducing the ventricular afterload. In conclusion, these results indicate that, from the perspective of cardiovascular physiology, a pulsatile LVAD with counter-pulsation operation is a plausible alternative to the existing LVAD with continuous flow mode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.