Abstract

Equivalent (effective) values of airway resistance were determined for total breaths, inspiration and expiration in spontaneously breathing patients by means of an off-line processing system. An improvement in the alveolar pressure determination was achieved by subtracting thermal pressure drifts of the pressure plethysmograph and by taking into account instantaneous values of the intrathoracic gas volume. Equivalent resistance values were compared with the results of several simulated routine evaluations. It was found that the equivalent resistance is best approximated by the graphical procedure using average pressure points at +/- 0.5 liter/sec flow levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.