Abstract

A method of computing theoretical X-ray spectra in the range 30-150 kV is presented. The theoretical spectra are compared with constant potential, high resolution spectra from a tungsten target measured with a Ge(Li) detector, for a range of target angles, tube voltage and filtrations. Above 100 kV the spectra were also measured with a NaI detector but, as there was good agreement between the Ge(Li) and NaI detectors, only the former are presented. Spectra computed using Kramers' theory are also included for comparison, giving fairly good agreement at large target angles (30 degrees ) but becoming gradually worse as the target angle decreased. Spectra may be computed by this method for any desired filtration, target angle, and tube voltage between 30 and 150 kV, in excellent agreement with the measured data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.