Abstract

Compton scattering of gamma rays within the image volume has been assessed for a large-aperture positron-emission-tomography imaging system. The Compton scattered attenuation and the Compton scattered background were both modeled and measured for point sources centered in scattering spheres up to 10 cm in diameter. Good agreement was obtained between simulations and measurements. The attenuation problem is independent of the detector system, but its correction is more difficult in a large-aperture system. The scattered coincidence background is large in this system (43% for a 10-cm-diameter scattering sphere), but the background overlap is reduced with 3D imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.