Abstract The design and construction of a gas Cerenkov-counter of one m 2 cross-sectional area and operating at pressure of up to 8 bar is described. The light collection is effected by multiple reflection on diffusely reflecting surfaces. Measurements of the detection efficiency show the counter to be 98% efficient for pions of 3.5 GeV/ c . The results obtained indicate that the construction of considerably larger detectors of this kind can be envisaged.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.