Abstract

This study aimed to research the effect of different transgingival heights (THs) and types of abutments on performances including the compressive resistances and fracture modes of abutments on internal conical connected implants. Three groups were established: a 1-piece zirconia abutment group (group A), a 2-piece zirconia abutment group (group B), and a 1-piece titanium abutment group (group C). Three THs (2, 3.5, and 5 mm, n = 6) were set as the subgroups of each group. All groups were subjected to a compressive resistance test at a 30 degree angle. The main failure modes were analyzed using scanning electron microscopy and a stereomicroscope. The compressive resistances of the abutments on the internal conical connected implants were significantly related to the TH (P = 0.004), the type of abutment (P < 0.0001), and the combined effect (P < 0.0001). All of the subgroups in group C exhibited the greatest compressive loads at the same TH. Different types of abutments had different failure modes. Both the TH and the type of abutments influenced compressive resistances of the implant-abutment complexes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.