Abstract

Compressed fluorescence lifetime imaging (Compressed-FLIM) is a novel Snapshot compressive imaging (SCI) method for single-shot widefield FLIM. This approach has the advantages of high temporal resolution and deep frame sequences, allowing for the analysis of FLIM signals that follow complex decay models. However, the precision of Compressed-FLIM is limited by reconstruction algorithms. To improve the reconstruction accuracy of Compressed-FLIM in dealing with large-scale FLIM problem, we developed a more effective combined prior model 3DTGp V_net, based on the Plug and Play (PnP) framework. Extensive numerical simulations indicate the proposed method eliminates reconstruction artifacts caused by the Deep denoiser networks. Moreover, it improves the reconstructed accuracy by around 4dB (peak signal-to-noise ratio; PSNR) over the state-of-the-art TV+FFDNet in test data sets. We conducted the single-shot FLIM experiment with different Rhodamine reagents and the results show that in practice, the proposed algorithm has promising reconstruction performance and more negligible lifetime bias.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.