Abstract

A coupled supercritical fluid extraction (SFE), high-speed gas chromatography (GC), and element-selective radiofrequency plasma detection (RPD) system was evaluated for the analysis of volatile organic compounds in soil matrices. The effect of GC carrier gas (carbon dioxide) and plasma reagent gas (oxygen) on the selectivity and sensitivity of the RPD was studied to establish the optimal operational conditions. The selectivities obtained were 3500 (Cl/C) for chlorine and 2600 (S/C) for sulfur. The detection limits were 24.8 and 9.2 pg s -1 for chlorine and sulfur, respectively. Complete quantitative analysis could be easily accomplished within 10 min

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.