Abstract

The PANDA (Particles And Non-Destructive Analysis) device is a measuring system developed for non-destructive analysis of samples for safety, security and safeguards. It consists of two complementary measurement positions. The position sensitive alpha–gamma coincidence setup of PANDA has been updated with a pile-up rejection circuit. This article demonstrates, using a 241Am source, that this circuit can be used to identify gamma-ray sum peaks in the spectra. The second measurement position reported here hosts a 10-mm2 windowless silicon drift detector that detects electrons and X-rays with a very good energy resolution. An experimental value of 0.30keV for full width at half maximum was obtained at 17.7keV energy (Np X-rays). Conversion electron spectrometry, for example, provides an independent, complementary means to study actinides whose low energy transitions are often highly converted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.