Abstract

The PANDA (Particles And Non-Destructive Analysis) device is a measuring system developed for non-destructive analysis of samples for safety, security and safeguards. It consists of two complementary measurement positions. The position sensitive alpha–gamma coincidence setup of PANDA has been updated with a pile-up rejection circuit. This article demonstrates, using a 241Am source, that this circuit can be used to identify gamma-ray sum peaks in the spectra. The second measurement position reported here hosts a 10-mm2 windowless silicon drift detector that detects electrons and X-rays with a very good energy resolution. An experimental value of 0.30keV for full width at half maximum was obtained at 17.7keV energy (Np X-rays). Conversion electron spectrometry, for example, provides an independent, complementary means to study actinides whose low energy transitions are often highly converted.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.