Abstract
ABSTRACT Federal Communications Commission (FCC) has released the regulation for medical body area network (MBAN) in 2012 as a BAN specialized for medical applications. MBAN is a narrowband BAN allocated on the 2360-2400 MHz frequency band. It is very promising, as it has the advantages of BAN modules like low power and high accuracy, which are essential parameters for medical applications. This paper provides a comprehensive study of the on-body MBAN operating in the 2360-2390 MHz band. In particular, it introduces link budget signaling parameters required for reliable transmission of biological signals for the MBAN channels, and investigates the bit-error-rate (BER) and maximum allowable data throughput performances assuming different node positions and body activities. Numerical results show that a single arm receiver is ideal for medical applications that do not require movement. Moreover, the relative movement between the transmitter and receiver nodes highly affects the performance of the link.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.