Abstract
We describe a compound parabolic concentrator (CPC)-based probe for enhanced signal collection in the spectroscopy of biological tissues. Theoretical considerations governing signal enhancement compared with conventional collection methods are given. A ray-tracing program was used to analyze the throughput of CPC's with shape deviations and surface imperfections. A modified CPC shape with 99% throughput was discovered. A 4.4-mm-long CPC was manufactured and incorporated into an optical fiber-based near-infrared Raman spectrometer system. For human tissue samples, light collection was enhanced by a factor of 7 compared with collection with 0.29-NA optical fibers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.