Abstract
The use of computer navigation during hip resurfacing has been proposed to reduce the risk of a malaligned component and notching with subsequent postoperative femoral neck fracture. Femoral component malalignment and notching have been identified as the major factors associated with femoral neck fracture after hip resurfacing. We performed 37 hip resurfacing procedures using an imageless computer navigation system. Preoperatively, we generated a patient-specific computer model of the proximal femur and planned a target angle for placement of the femoral component in the coronal plane. The mean navigation angle after implantation (135.5 degrees) correlated with the target stem-shaft angle (135.4 degrees). After implantation, the mean stem-shaft angle of the femoral component measured by three-dimensional computed tomography (135.1 degrees) correlated with the navigation target stem-shaft angle (135.4 degrees). The computer navigation system generates a reliable model of the proximal femur. It allows accurate placement of the femoral component and provides precise measurement of implant alignment during hip resurfacing, thereby reducing the risk of component malpositioning and femoral neck notching.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.