Abstract

Microwave hyperthermia applicators are generally characterized at two or three frequencies. These frequencies usually are at the extremes of the applicator's operating range. However, such applicators are often used at arbitrary frequencies within their range. The most common reason for choice of frequency is to achieve acceptable coupling in the clinical configuration with a given patient. Occasionally a spurious transmission pattern will result from the frequency chosen that can lead to undetected high power densities well away from the target volume. In the present report, such a situation is discussed. The transmission pattern resembled “horns” reported at the field edge of certain accelerators. This pattern resulted in the inability to heat the target volume and the production of thermal blisters outside that volume. Such an occurrence underscores the need to characterize the transmission characteristics of a microwave hyperthermia applicator at all the frequencies used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.