Abstract
The sixth annual survey of complications associated with foldable intraocular lenses (IOLs) requiring explantation or secondary intervention was sent to members of the American Society of Cataract and Refractive Surgery and the European Society of Cataract and Refractive Surgeons. Preoperative data about visual acuity, patient signs and symptoms, and complications requiring IOL removal were evaluated. Complications were then tabulated for each of the following major foldable IOL groups: 3-piece monofocal silicone, 3-piece hydrophobic acrylic, 1-piece hydrophobic acrylic with haptics, 3-piece hydrophilic acrylic (hydrogel), 1-piece hydrophilic acrylic (hydrogel), 1-piece plate-type silicone, 3-piece multifocal silicone, and Collamer. Two hundred seventy-three surveys were returned for evaluation. Dislocation/decentration, incorrect lens power, IOL calcification, and glare/optical aberrations were the most common reasons for removing foldable IOLs. Good surgical technique, accurate IOL power measurements, and high manufacturing standards for foldable IOL materials and designs are the most important factors in avoiding complications with foldable IOLs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.