Abstract

To analyze complications of femtosecond lasers used for cataract surgery. Department of Ophthalmology Semmelweis University, Budapest, Hungary. Retrospective analysis. Intraoperative complications of the first 100 femtosecond laser-assisted (Alcon-Lensx, Inc.) cataract surgeries were collected. Possible complications of femtosecond capsulotomies and their management were also assessed. The complications were as follows: suction break (2%), conjunctival redness or hemorrhage (34%), capsule tags and bridges (20%), anterior tear (4%), miosis (32%), and endothelial damage due to cut within the endothelial layer (3%). There were no cases of capsule blockage or posterior capsule tear. During the learning curve, there was no complication that would require vitrectomy. All complications occurred during the first 100 cases. Femtosecond laser cataract surgery had a learning curve during the first 100 cases. With cautious surgical technique, the complications can be avoided. The femtosecond laser-assisted method was efficient and safe for cataract surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.