Abstract

A retrospective cross-sectional analysis was conducted of the US Food and Drug Administration's MAUDE (Manufacturer and User Facility Device Experience) database, to evaluate the complication profile of cochlear implantation according to manufacturer. A review of the MAUDE database was conducted from 1 January 2010 to 31 December 2020. Complications, including infection, extrusion, facial nerve stimulation, meningitis and cerebrospinal fluid leak, were identified using key word searches. The categorised data were analysed using a chi-square test to determine a difference in global complication incidence between three major cochlear implant manufacturers: manufacturer A (Cochlear Limited), manufacturer B (Med-El) and manufacturer C (Advanced Bionics). A total of 31 857 adverse events were analysed. Implants of manufacturer C were associated with a statistically higher rate of infection (0.97 per cent), cerebrospinal fluid leak (0.07 per cent), extrusion (0.44 per cent) and facial nerve stimulation (0.11 per cent). Implants of manufacturer B were associated with a statistically higher rate of meningitis (0.07 per cent). Consideration of patient risk factors along with cochlear implant manufacturers can heighten awareness of cochlear implant complications pre-operatively, intra-operatively and post-operatively.