Abstract

From a group of 643 total hip arthroplasties, 98 patients required trochanteric osteotomies, and in 68 the trochanteric osteotomies were repaired with the Dall-Miles cable grip system (Howmedica, Rutherford, NJ). Trochanteric nonunion occurred in 17 of these patients (25%), with fraying and fragmentation of the cable present in 15 (88%). Of the 51 patients with radiographic union, 18 (35%) also had signs of fraying and fragmentation. Bone destruction around the cable in the area of the lesser trochanter was seen in seven patients (10%). Large deposits of metal debris at the inferior border of the acetabulum were seen in eight hips (12%). Multifilament cable did not appear to offer significant advantages over standard monofilament wire, and the potential problems of fraying, fragmentation, and free-floating metallic debris must be considered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.