Abstract

Glucose monitoring is an integral part of diabetes management. Interstitial glucose monitoring systems are increasingly commonly being used for this purpose in dogs and cats, including the use of a flash glucose monitoring system (FGMS). The aim of this study was to describe the incidence and nature of complications associated with the use of an FGMS in diabetic cats. The medical records of all cats that had placement of a 14-day FGMS during a 1-year period were retrospectively reviewed. Data retrieved included the number of days the sensor remained attached and functional, location of sensor placement and complications associated with the sensor. Complications were defined as early sensor detachment, sensor failure prior to the end of the 14-day monitoring period and dermatologic changes at the sensor site. Descriptive statistics were used to characterize the data. Twenty cats had a total of 33 FGMSs placed. The majority (30/33 [91%]) of sensors were placed over the dorsolateral aspect of the thorax just caudal to the scapula. Twenty (61%) FGMSs remained attached and functional for the full 14 days. The overall incidence of complications associated with FGMS use was 10/33 (30%). The most frequent complication was early sensor detachment (n = 5/33 [15%]). Mild dermatologic changes (erythema, crusts) were noted with 4/33 (12%) FGMSs. More serious complications (skin erosions, abscess formation) were noted with 2/33 (6%) FGMSs. The use of the FGMS is relatively safe in cats, although there are potential complications that owners should be made aware of.