Abstract
Although the manufacturers' specifications may be identical, brands of widely employed straight, long-paper angiographic guidewires vary considerably in performance. Because guidewire flexibility is a key factor in performance, we investigated the characteristic compliance patterns of similar guidewires (using an apparatus of our own design) from four different manufacturers. The brands were found to differ significantly in compliance at the tip, in rate of change in compliance, and in overall length of taper.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.