Abstract

A beam deflection time-of-flight mass spectrometer was developed in conjunction with an integrating transient recorder to provide time array detection, permitting high mass spectral scan file acquisition rates for complex mixture analysis by capillary gas chromatography-mass spectrometry (GC-MS). Results are presented for the analysis of a urinary organic acid mixture by GC-MS at a scan file acquisition rate of 10 scan files per second (sf/s), showing the advantages of such data collection in the deconvolution of partially resolved components. The reconstructed total ion current (RTIC) chromatogram available from data acquired at this scan file generation rate is shown to be comparable to the profile obtained from a flame ionization detector in representing the chromatography performed under identical experimental parameters. The RTIC chromatogram available from the database obtained at 10 sf/s is compared with that available from a database obtained at 1 sf/s, the latter representing that scan rate typically used with most GC-MS instruments. The advantages of the higher scan file acquisition rate in representing the chromatographic profile and in allowing mass spectral data to be obtained for components in the complex mixture that are unresolved chromatographically are discussed.