Transient reproducible complete infrahisian atrioventricular block occurred after endocavitary cardioversion of ventricular tachycardia and ventricular fibrillation by an automatic implantable cardiac defibrillator in a patient with left bundle branch block and baseline borderline H-V interval. Six months later, a permanent complete atrioventricular block occurred after repetitive endocavitary shocks. Several hypothetical mechanisms may explain this infrahisian atrioventricular block: a “fatigue” phenomenon of the His-Purkinje system, vagally mediated atrioventricular block, and/or ablation of the His conduction system. The last seems to be the most probable, which would justify implantation of an automatic implantable cardiac defibrillator with backup pacing in all patients, especially in the presence of bundle branch block.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.