Abstract
We describe an uncommon and potentially life-threatening complication during stent assessment by optical coherence tomography (OCT) with entrapment and complete stent removal. Possible mechanisms for OCT-catheter entrapment include malapposed stent struts, catheter deformation, and loss of guidewire position. In this case, the OCT catheter was advanced over the guidewire, which was probably passed through a proximal malapposed stent strut. Tips to avoid potential complications during OCT performance are provided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.