Abstract

A systematic evaluation of activity concentrations of 3H, 63Ni and 14C by using the CIEMAT/NIST method has been carried out over time intervals of 4, 2.4 and 2.3 years, respectively. It proves that the CIEMAT/NIST method can compensate, with no need for additional checking sources or procedures, for count losses and quench parameter reduction in long-term measurements with LSC spectrometers and allows one to obtain very accurate activity concentration values by using a quench curve close to the nuclide measurements. If this requirement of the method is not fulfilled, significant discrepancies, up to 5.6% for 3H, 1.5% for 63Ni and 0.35% for 14C, can be obtained in only 1-year time difference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.