Abstract

To compare corneal electrodes commonly used in rodent eyes for repeat and left versus right eye accuracy and variability to record the flash electroretinogram (ERG). Animals studied were eight C57BL/6 mice and eight rats of the Wistar strain. Scotopic ERGs were recorded from eyes of dark-adapted anesthetized rodents to compare a custom-made gold-wire contact lens electrode (CLE), a cotton-wick silver-silver chloride electrode (SCLE), and a coiled stainless steel wire electrode (SSE). Compared to SCLE and SSE. the potentials recorded by CLE are characterized by significantly larger ERG amplitudes and oscillatory potentials in both rats and mice (p <0.0001). In analyzing test-retest data comparing the three different electrodes the coefficient of variation was smaller (range, 10.3-15.5%) and the interclass correlation coefficient (0.77-0.93) showed a better agreement for the CLE. Recording scotopic ERGs with custom-made gold-wire contact lens electrodes records large amplitudes and shows a good reproducibility and reliability to monitor retinal function in rodent eyes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.