Abstract

To compare intraocular pressure (IOP) readings between Tono-Pen tonometry and GAT, between noncontact tonometry (NCT) and GAT, and between dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT). The correlation between IOP reading and possible confounder was identified. This observational cross-sectional study enrolled sixty-two healthy subjects. All IOP and ocular pulse amplitude (OPA) measurements were taken by a single ophthalmologist; mean keratometric power (MK), central corneal thickness (CCT), and lens thickness (LT) were measured by a single experienced technician. Stepwise multiple regression analysis indicated that GAT (P=0.017) and DCT (P=0.002) readings correlated positively with MK; GAT, NCT, and Tono-Pen readings correlated positively with CCT (P<0.05); NCT (P=0.035), and DCT (P=0.016) readings correlated negatively with LT; GAT (P=0.006) and Tono-Pen (P=0.009) readings correlated positively with OPA. The K, CCT, LT, and OPA are confounders in tonometry readings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.