Abstract

Background and ObjectivesA rotation chair test has been used to evaluate the function of the horizontal semicircular canals. Currently, two chair systems according to the presence of cylindrical darkroom are used in a clinic setting. However, it has not been thoroughly investigated whether one system is superior to the other system or not. In this study, we aimed to compare test outcomes and subject convenience between two systems.Subjects and MethodsTwenty subjects with no history of otologic disease were enrolled. Subjects were tested with two systems: system [A] with a cylindrical chamber and system [B] with no chamber. The results of sinusoidal harmonic acceleration (SHA), step velocity (SV), and visual fixation (VFX) tests were compared between the systems. Subject convenience was assessed with a questionnaire survey and results were compared between the systems.ResultsThere were no significant differences in gain or asymmetry in SHA test between the systems. However, the phase of system [A] was significantly lower than that of system [B] at 0.16 Hz. There was no significant difference between the systems in directional preponderance (DP) gain or DP time constant. Regarding the VFX test, gain was higher in system [A] than system [B]. Subjects reported less stuffiness and less anxiety with system [B] than system [A], while preferring the system [A] goggles.ConclusionsA rotation chair system without a darkroom can provide a more comfortable experience for subjects in terms of stuffiness and anxiety, while showing comparable results in SHA and SV tests with a darkroom system.