Abstract
The present study was realized for measuring viscosities of reference liquids using capillary viscometers and Stabinger viscometer SVM 3001 with viscosity interval between 1 mm2/s and 5000 mm2/s with temperatures from 20 °C to 80 °C. Based on our measurement with various liquids, we determine the viscosity values and compare both of the results. The aim of this study to evaluate the results of the primary level viscosity measurement system and stabinger viscometer and to compare the measurement results due to the providing traceability of Stabinger viscometer by TUBITAK UME. An increasing number of national metrology institutes and accredited laboratories provide viscometer calibration with reference liquids in a wide viscosity range. It is a common practice to use the viscosity of water as the metrological basic of viscometry. The national standard of viscosity provided by TUBITAK UME consists of a set of ubbelohde viscometers covering the measuring range of kinematic viscosities from about 0.5 mm2/s to 100 000 mm2/s. At the low viscosity, long − capillary viscometers are used as primary standards which are directly calibrated water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Metrology and Quality Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.