Abstract

Indirect photometric and unsuppressed direct conductivity detection modes are examined using naphthalene mono-, di-, and tri-sulfonate as mobile phases for the separation of several anions such as F−, Cl−, NO2−, Br−, NO3−, SO4=,I−, and SCN− using a commercial anion exchange column. With all three mobile phases, conductivity detection shows better sensitivities and detection limits than indirect photometry. Conductivity detection is 5 to 16 times more sensitive than indirect photometry for all analytes. Detection limits achieved using these mobile phases are, for example, 0.04 ng and 0.1 ng for chloride ion with conductivity and indirect photometry, respectively. Both detection modes give wide linear ranges extending from at least 100 ppm to the detection limit of each anion which is generally about 0.02 ppm. Sulfur oxide anions such as dithionate and tetrathionate are separated using flow programming with naphthalenetrisulfonate as the mobile phase in less than 20 minutes. With both detection modes, desired chromatographic performance of these three eluents is achieved without pH adjustment of the mobile phase.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.