Abstract

Absolute measurement of radiant power in the X-ray region is essential for many applications in astrophysics, spectroscopy, and X-ray diagnostics. Comparison between different measuring methods is an effective way to check their reliability. In the present work, a comparison of X-ray radiant power absolute measurement between a free-air ionization chamber and a cryogenic electrical substitution radiometer was performed at Beijing Synchrotron Radiation Facility. The absolute radiant power obtained by these two methods were mutually compared via a transfer standard detector's spectral responsivity at a photon energy of 10 keV. The result of the comparison showed that the difference was 0.47%. A conclusion was reached that the free-air ionization chamber and the cryogenic electrical substitution radiometer agreed within the combined relative uncertainty of 3.35%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.