Abstract
BackgroundPublished treatment technique comparisons for postoperative left-sided whole breast irradiation (WBI) with deep-inspiration breath-hold (DIBH) are scarce, small, and inconclusive. In this study, fully automated multi-criterial plan optimization, generating a single high-quality, Pareto-optimal plan per patient and treatment technique, was used to compare for a large patient cohort 1) intensity modulated radiotherapy (IMRT) with two tangential fields and 2) volumetric modulated arc therapy (VMAT) with two small tangential subarcs.Materials and methodsForty-eight randomly selected patients recently treated with DIBH and 16 × 2.66 Gy were included. The optimizer was configured for the clinical planning protocol. Comparisons between IMRT and VMAT included dosimetric plan parameters, estimated excess relative risks (ERR) for toxicities, delivery times, MUs, and deliverability accuracy at a linac.ResultsThe automatically generated IMRT and VMAT plans applied in this study were similar or higher in quality than the manually generated clinical plans. For equal PTVin V95% (98.4 ± 0.9%), VMAT had significant advantages compared to IMRT regarding breast dose homogeneity and doses in heart and ipsilateral lung, at the cost of some minor deteriorations for contralateral breast (few cases with larger deteriorations) and lung. Conformality improved from 1.38 to 1.18 (p < 0.001). With VMAT, ERR for major coronary events and ipsilateral lung tumors were reduced by 3% (range: −1–12%) and 16% (range: −3–38%), respectively. MUs and delivery times were higher for VMAT. There were no statistical differences in γ passing rates.ConclusionFor WBI in conservative therapy of left-sided breast patients treated with DIBH, VMAT with two tangential subarcs was generally dosimetrically superior to IMRT with two tangential static fields. Results need confirmation by robustness analyses.
Highlights
Published treatment technique comparisons for postoperative left-sided whole breast irradiation (WBI) with deep-inspiration breath-hold (DIBH) are scarce, small, and inconclusive
Normalization of autoIMRT and autoVMAT plans to obtain the same PTVin coverage as the corresponding clinical plans resulted in an average PTVin V95% of 98.4 ± 0.9%, range 94.9–99.7
Higher in quality than the clinical plans (Fig. 1). This observation demonstrates that the comparisons in this study between Intensity-modulated radiotherapy (IMRT) and volumetric arc therapy (VMAT) for WBI with DIBH were based on treatment plans of sufficiently high dosimetric quality
Summary
Published treatment technique comparisons for postoperative left-sided whole breast irradiation (WBI) with deep-inspiration breath-hold (DIBH) are scarce, small, and inconclusive. Fully automated multi-criterial plan optimization, generating a single high-quality, Pareto-optimal plan per patient and treatment technique, was used to compare for a large patient cohort 1) intensity modulated radiotherapy (IMRT) with two tangential fields and 2) volumetric modulated arc therapy (VMAT) with two small tangential subarcs. Comparisons between IMRT and VMAT included dosimetric plan parameters, estimated excess relative risks (ERR) for toxicities, delivery times, MUs, and deliverability accuracy at a linac. Intensity-modulated radiotherapy (IMRT) with static beams, and later volumetric arc therapy (VMAT), have been proposed to improve breast dose homogeneity and possibly reduce dose to K. For VMAT, the two static tangential IMRT fields are often replaced by two small tangential arcs or by a single, larger partial arc [5, 6]
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.