Abstract
Various electrode materials commonly used in electroanalysis were tested in preparation of respective bismuth film electrodes and compared in simultaneous electrochemical detection of heavy metals, simulating dissolved quantum dots used as labels of antibodies in quantum dot-linked immunosorbent assays. According to results obtained during voltammetric measurements, it can be concluded that selected commercial screen-printed electrodes showed many significant drawbacks which are described in detail. Contrary to the gold disk electrode, conventional glassy carbon electrode coated with thin bismuth film was applicable up to −1.5 V without the need to use very high content of bismuth in supporting electrolyte. Moreover, significant interactions among present metals were observed during their electrochemical stripping analysis, especially in presence of copper and silver.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.