Abstract
Osmometry is the analytical method of measuring osmotic pressure (π) and related colligative properties. Accurately measuring π is essential for applications that use osmosis, including drug delivery systems and water filtration methods such as forward osmosis. The three most common forms of directly or indirectly measuring π are vapour pressure osmometry (VPO), freezing point osmometry (FPO), and direct membrane osmometry (DMO). In this report, the performance of VPO, FPO, and DMO are compared for various aqueous solutions, including small molecule and polymer solutes, and potentially more difficult samples such as concentrated, carbonated, or viscous solutions. Potential problems with each of the three methods are examined. Good practices can avoid inaccuracies in DMO due to external concentration polarization and solute diffusion through the membrane. Similarly, it is important to be aware of and avoid weaknesses in VPO and FPO in order to obtain reliable results. Even with care, however, the three methods still give different values of π for concentrated aqueous solutions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.