Abstract

This study aimed to develop a simple UV spectrophotometric method for the analysis and the dissolution test of flunarizine in capsules. The UV absorbance was both measured directly and by the first derivative measurements at 254 and 268 nm, respectively. The developed methods were validated for their linearity, accuracy, precision, limit of detection (LOD) and limit of quantitation (LOQ) in comparison with the reported HPLC method. The UV spectrophotometric method illustrated excellent linearity ( r 2 > 0.9999) in the concentration range of 6–24 μg/mL. Precision (%R.S.D. < 1.50) and recoveries were good (% R > 99.62). The LOD of direct UV and first derivative measurements were 0.09 and 0.84 μg/mL, respectively, and the LOQ were 0.26 and 2.55 μg/mL, respectively. Results from the assay of flunarizine in capsules by the UV spectrophotometric methods, both direct and first derivative measurements were not significantly different from those of the HPLC method ( P > 0.05). Additionally, the method was successfully used for the dissolution test of flunarizine capsule and was found to be reliable, simple, fast, and inexpensive.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.