Abstract
The neutron spectra and ambient dose equivalent were determined by unfolding measured Bonner sphere system data using different unfolding methods. These methods included a maximum entropy method (MAXED code), nonlinear least-squares method (GRAVEL code) with several different starting spectra, and a genetic algorithm method. These algorithms were used to unfold measured Bonner sphere data that had been collected using a LiI(Eu) detector and activation foils. The Bonner sphere system was exposed to neutrons from a known AmBe source and at the proton accelerator facility at the University of Texas M. D. Anderson Cancer Center. The total fluence rates and total ambient dose equivalents of the neutron field agree within 7.5%, regardless of unfolding algorithm or starting spectrum. In contrast, the fluence-weighted average energy varied dramatically, depending on the starting spectrum used in the unfolding process. These findings offer insight and guidance into the use of unfolding algorithms and starting spectra for neutron spectroscopy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.