Abstract

The determination of isotopic thorium by alpha spectrometric methods is a routine practice for bioassay and environmental measurement programs. Alpha-spectrometry has excellent detection limits (by mass) for all isotopes of thorium except 232Th due to its extremely long half-life. This paper discusses improvements in the detection limit and sensitivity over previously reported methods of pre-concentration neutron activation analysis (PCNAA) for the recovery corrected, isotopic determination of thorium in various matrices. Following irradiation, the samples weredissolved, 231Pa added as a tracer, and Pa isolated by two different methods and compared (extraction chromatography and anion exchange chromatography) followed by alpha spectrometry for recovery correction. Ion exchange chromatography was found to be superior for this application at this time, principally for reliability. The detection limit for 232Th of 3.5 · 10-7 Bq is almost three orders of magnitude lower than foralpha spectrometry using the PCRNAA method and one order of magnitude below previously reported PCNAA methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.