Abstract

In this study, we evaluated gas plasma surface sterilization methods in a specific sterilizer. We have introduced a new monitoring method using 0.4 microm pore size membranes, which in this study gave the information corresponding to 3000 exposed biological indicators per treatment cycle. This enabled us to compare the fraction of inoculates that showed no growth after exposure for 30 different locations in the chamber, and hereby identify weak and strong spots in the chamber with regard to sporicidal effect. Membranes were also used to expose a broad spectrum of soil bacteria for plasma treatment at four different conditions. The organisms were identified using PCR and sequencing. The test showed that Bacillus stearothermophilus spores were inactivated at the slowest rate among the tested microorganisms. Further alpha-proteobacteria (Gram negative) seemed more sensitive than the rest of the tested organisms. The microspot evaluation approach has been a most useful tool in the assessment of sterilization performance in sterilizers that do not have clear measurable parameters related to the sterilization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.