Abstract

Diagnosis of Human T-cell Lymphotropic Virus (HTLV) types I and II infection requires sequencial testing with firstly a screening using an Enzyme immunoassay followed by a confirmatory test. To compare the performances of the Alinity i rHTLV-I/II (Abbott®) and LIAISON® XL murex recHTLV-I/II serological screening tests to the ARCHITECT rHTLVI/II test followed if positive by HTLV BLOT 2.4, MP Diagnostics as the reference. 119 serum samples from 92 known HTLV-I infected patients and 184 from uninfected patients with HTLV were analyzed in parallel with, Alinity i rHTLV-I/II, LIAISON® XL murex recHTLV-I/II and ARCHITECT rHTLVI/II. Alinity i rHTLV-I/II and LIAISON® XL murex recHTLV-I/II exhibited a total agreement with ARCHITECT rHTLVI/II for both positive and negative samples. Both tests are suitable alternatives for HTLV screening.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.