Abstract

IntroductionThe aim of the study was a comparative evaluation of in-house real-time PCR and commercial real-time PCR (Fast Track Diagnostics (FTD), ampliCube/Mikrogen) targeting enteropathogenic bacteria from stool in preparation of Regulation (EU) 2017/746 on in vitro diagnostic medical devices.MethodsBoth 241 stool samples from patients and 100 samples from German laboratory control schemes (“Ringversuche”) were used to comparatively assess in-house real-time PCR, the FTD bacterial gastroenteritis kit, and the ampliCube gastrointestinal bacterial panels 1&2 either with the in-house PCRs as gold standard and as a test comparison without gold standard applying latent class analysis. Sensitivity, specificity, intra- and inter-assay variation and Cohen’s kappa were assessed.ResultsIn comparison with the gold standard, sensitivity was 75–100% for strongly positive samples, 20–100% for weakly positive samples, and specificity ranged from 96 to 100%. Latent class analysis suggested that sensitivity ranges from 81.2 to 100% and specificity from 58.5 to 100%. Cohen’s kappa varied between moderate and nearly perfect agreement, intra- and inter-assay variation was 1–3 to 1–4 Ct values.ConclusionAcceptable agreement and performance characteristics suggested replaceability of the in-house PCR assays by the commercial approaches.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.