This paper compares two complementary-metal-oxide semiconductor front-end transimpedance amplifiers (TIA) intended for use in optical biosensors. They are the shunt-feedback and current-mode circuits, the most widely used topologies for wideband operation. The former consists of a three-stage nested-Miller-compensated amplifier in noninverting mode with a photodiode bootstrapping and a controlled voltage gain; the latter comprises a wideband common-gate feedback current mirror coupled to a current-to-voltage conversion stage and two common-source gain stages. The post-layout simulation results show that the shunt-feedback TIA achieves a maximal gain of 112 dBΩ over a 2-MHz bandwidth, whereas the current-mode TIA has a flat gain of roughly 83 dBΩ over a 115-MHz bandwidth. The overall input rms noise of each circuit was 185 pA/√Hz and 53 nA/√Hz, respectively, with power consumptions of 0.5 and 28.6 mW. We find that the shunt-feedback TIA is a better choice for high-resolution low- to mid-frequency applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.