Abstract

To assess the agreement of intraocular pressure (IOP) measured with the Tono-Pen and the Goldmann applanation tonometer (GAT) in normal children and adolescents. A total of 439 subjects from birth to <18 years of age without anterior segment anomalies or glaucoma had their IOP measured with the two instruments by separate, masked examiners in the office or under general anesthesia. On average, the Tono-Pen measured values slightly lower than the GAT for IOP <11 mm Hg and slightly higher than the GAT for IOP >11 mm Hg in the office setting. Using the average of GAT and Tono-Pen IOPs to estimate the true IOP, the average difference (GAT - Tono-Pen) was 0.4 mm Hg at IOP of 10 mm Hg and -3.0 mm Hg at IOP of 20 mm Hg. The 95% limits of agreement on the average difference between instruments were ± 6.4 mm Hg in the office setting and ± 6.8 mm Hg under general anesthesia. Larger differences between instruments were found with younger age. Standard error of measurement with the Tono-Pen was 1.44 mm Hg and 1.82 mm Hg for the office and anesthesia settings, respectively. Thicker corneas were associated with higher IOP with both the GAT and the Tono-Pen. In normal children, average differences between IOP measured by Tono-Pen and GAT were small, although there was substantial test-retest variability. Younger age was associated with larger average differences, as was higher IOP in the office setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.