Purpose: To compare retinal nerve fiber layer (RNFL) thickness obtained with Stratus optical coherence tomography (OCT) and Cirrus OCT. Methods: Sixty-one normal eyes were evaluated with Stratus and Cirrus OCT on the same day, and the RNFL thicknesses measured by the two OCT machines were compared. The correlation between the two data sets was obtained using Pearson's correlation coefficient. The correlation between RNFL thickness and the difference in data measured by the two OCT machines was then assessed. Results: The average RNFL thickness was significantly higher with Stratus OCT by 6.54 (p=0.0008). A strong correlation was present between the two RNFL thickness data sets (r=0.883), and the difference between Stratus and Cirrus values tended to increase as RNFL thickness increased. Conclusions: RNFL thickness measurements in normal eyes scanned with Cirrus OCT correlate well with Stratus OCT measurements. Average RNFL thickness was significantly higher with Stratus OCT, and as the RNFL thickness increased, the difference between Stratus and Cirrus values increased.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.