Abstract

Radioimmunoassays for methotrexate are described, involving use of a rabbit antiserum to a conjugate of the drug and bovine serum albumin and the drug labeled with tritium, selenium-75, or iodine-125. Of the two gamma emitters, the 75Se-labeled drug was prepared by the Radiochemical Centre, Amersham, England and the 125I-labeled drug in the laboratory, by the Chloramine T iodination technique. The stability of labels with both methods allows use of the faster, cheaper, and simpler gamma-counting techniques, with results available after 3 h. All three methods have acceptable sensitivity, accuracy, precision, and reproducibility, and are specific for methotrexate, with no significant interference from naturally occurring folates or leucovorin. The assays in which the gamma emitters are used have significant practical advantages over the beta emitter and are much better suited to automation and clinical application. The main advantage of 75Se=labeled methotrexate is its longer half-life, 121 days, as compared with 60 days for 125I.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.