Abstract

To compare the performance of 3 point-of-care glucose meters in adult and juvenile alpacas with that of a laboratory-based analyzer. Evaluation study. 35 adult alpacas and 21 juvenile alpacas. Whole blood samples obtained via jugular venipuncture were tested with all 3 point-of-care glucose meters; plasma samples were also tested with 1 of those meters. Glucose concentrations determined by use of the point-of-care meters were compared with results from the laboratory-based analyzer. Plasma glucose concentrations determined by use of the laboratory-based analyzer ranged from 36 to 693 mg/dL. Over the entire range of glucose concentrations tested, the Lin concordance correlation coefficient (agreement) was significant and excellent for all comparisons. Concordance decreased for 1 glucometer when testing whole blood samples over a narrower range of glucose concentrations (50 to 200 mg/dL). Bias was typically small (< 10 mg/dL) for 3 of the 4 comparisons but considerable for 1 meter with the use of whole blood. The limits of agreement were wide for all comparisons over the entire range of glucose concentrations tested but decreased to within acceptable limits when the narrower glucose range (50 to 200 mg/dL) was analyzed for 3 of the comparisons. For samples with a PCV < 25%, bias and the limits of agreement were greater for one of the meters tested. Discrepancies between point-of-care glucose meters and reference techniques can be considerable in alpacas, emphasizing the importance of assessing individual meter performance in a target population.