Abstract
We evaluated three commercial colistin susceptibility testing methods using 213 bloodstream Acinetobacter isolates identified by gene sequencing. Compared to the agar dilution reference method, excellent categorical agreements (both 99.1%) were observed using Vitek 2 and Etest, compared to 87.3% (95.7% for Acinetobacter baumannii and 80.7% for non-baumannii Acinetobacter isolates) using MicroScan.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.