Abstract

A direct comparison of the standards for air kerma of the Korea Research Institute of Standards and Science (KRISS), Korea, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60Co radiation beam of the BIPM in 2010. The result, expressed as a ratio of the KRISS and the BIPM standards for air kerma, is 0.9995 with a combined standard uncertainty of 1.6 × 10−3.The degrees of equivalence between the KRISS and the other participants in the key comparison, designated as BIPM.RI(I)-K1, have been calculated and the results are presented in the form of a matrix. A graphical presentation is also given.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).