Abstract

A comparison of the standards for absorbed dose to water of the National Metrology Institute of Japan (NMIJ) and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60Co γ-ray reference beam of the BIPM in November to December 2009 under the auspices of the key comparison BIPM.RI(I)-K4. The comparison result, based on the calibration coefficients for two transfer standards and expressed as a ratio of the NMIJ and the BIPM standards for absorbed dose to water, is 0.9960(46). The degrees of equivalence between the NMIJ and the other participants in this comparison have been calculated and the results are given in the form of a matrix for the ten national metrology institutes (NMIs) that have participated. A graphical presentation is also given. Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.